Brinell Hardness Testing Machine

Computerized Brinell Hardness Testing Machine - Model M B3000 PC

Features

Basic machine design and operation similar to B-3000(H). In addition, on 'Optical Device' with 14x magnification provided in front to project dia of ball impression on glass screen with a micrometer measuring system with 0.01 mm accuracy. The indentor swivels and project dia of ball impression immediately after unloading operation witch avoids additional time for measurement of ball impression. This given real Brinell number on production testing.

Available Models :

This model is available in 3 models.

  • Regular Computerised Machine.
  • Semi-automatic operation
  • Fully Automatic Machine means loading / unloading and results will get automatically.
The software includes:
  • Facility for Auto / semi Auto / Manual of operation.
  • Well managed database saves readings, w.r.t. batch.
  • Facility for calibration and check of calibration.
Specification:

The range of Measurement is from 1 mm to 6 mm of diameter with Resolution of 0.01 mm


Specification

SPECIFICATION
UNIT
M B3000 PC
Loads kg. 500 to 3000 in stages of 250
Initial Load kgf. Nill
Max. Testing Height mm 380
Depth of Throat mm 200
Max. Depth of elevation screw below base mm 180
Size of Basic mm 370 x 670
Machine Height mm 1185
Net Weight (Approx.) kg. 500
Drive Motor HP 0.33 - 415v/p

Applications
  • To measure Brinell Hardness directly on machine where presently Brinell-Microscope is used. This avoids eye straining of operator on producing testing. In addition it given for better repeated accuracy. High-low limits selection enables operator easy acceptance / rejection of components.
  • Can be connected to Loptop or Desktop PC, Easy to carry anywhere and handy for use with easy setup.
Scope of Supply:
  • Machine containing CCD Camera, optics and illumination systems with connecting cable.
  • External Vidio capture with USB 2 connectivity and driver software.
  • Brinell impression Measurement System (B.l.M.S.) software.
  • PC or Laptop and window XP operating system is to be procured by customer
  • One USB 2.0 and one Rs232 (serial) interface (9 pin) is required and it is recommended for Laptop to use external USB to serial converter for Rs232 interface.

Calibration Facility :

The systems includes calibration and check facilities to calibrate the systems and for checking of calibration. This eliminates any system error in measurement.